Osmomat 050
Technical Facts
|

|
| |
| measurement volume |
< 10 µL |
| sample size |
> 100 µL |
| measurement time |
1-3 Minutes, Depending on membrane type |
| measurement format |
99.9 mm Hg,
9.99 k-Pascal |
| reproducibility |
<± 0.5 mm Hg |
| significant figures |
3 places |
| zero point determination |
Automatic |
membrane
material |
Two layered membrane, each layer is able to retain compounds with molecular weight > 10,000 - 30,000 Dalton |
| recorder Output |
2mV/Digit, ±10%, can be set for recorder compensation with 1 Volt output |
| measurement |
Direct pressure measurement via difference in hydrostatic pressure |
| automation |
1. Stand alone wash of the upper half of the cell is available to prepare the cell
2. Storage of the results
3. Storage of all data values over a specified time period after the first measurement
|
|
© 2004 gonotec GmbH - Impressum
|